Infusion Device Analyzer

Safety Information

3

Avoid possible contamination of reusable components due

to backflow conditions. Some older style infusion devices
may have reusable components that could come in direct
contact with the fluids being pumped. When testing these
types of devices take care to avoid possible contamination
of reusable components.

Do not use delivery set or components that have been used

for testing for patient infusion.

Do not connect the Product to a patient or equipment

connected to a patient. The Product is intended for

equipment evaluation only and should never be used in

diagnostics, treatment or in any other capacity where the

Product would come in contact with a patient.

The Product must be properly earthed. Only use a supply

socket that has a protective earth contact. If there is any

doubt as to the effectiveness of the supply socket earth, do

not connect the Product. Do not use a two-conductor

adapter or extension cord. This will break the protective

ground connection.

Many components on the printed circuit board are static

sensitive. ESD precautions should be observed when

handling the printed circuit board assembly.

To avoid shock hazard and for proper Product operation,

connect the factory supplied three-conductor line power

cord to a properly grounded power outlet. Do not use a two-

conductor adapter or extension cord; this will break the

protective ground connection.

The Product is intended for use by trained service

technicians to perform periodic inspections on a wide range

of medical equipment. The testing procedures are menu-

driven and simple to operate.

The Product is intended for use with single-phase, grounded

power. It is not intended for dual, split-phase or three-phase

power configurations. But it can be used with any power

system that supplies the correct voltages for single-phase

and is grounded.

This Product is intended to be used in the laboratory

environment, outside of the patient care area, and is not

intended for use on patients, or to test devices while

connected to patients. This Product is not intended to be

used to calibrate medical equipment and not intended for

over the counter use.