Infusion Device Analyzer

General Specifications

33

General Specifications

Operating Voltage Range ..................................... 100 V ac to 240 V ac
Supply Frequency ................................................. 50/60 Hz
Supply Power ........................................................ <50 VA
Fuses ...................................................................... 20 mm T1.6 A H 250 V x 2
Size (H x W x D) ..................................................... 30 cm x 20 cm x 20 cm (12 in x 8 in x 8 in)
Weight .................................................................... ~5 kg (11 lb)
Altitude ................................................................... 0 m to 3000 m (10000 ft)
Temperature

Operating ............................................................ 15

°

C to 30

°

C (59

°

F to 86

°

F)

Storage ............................................................... -20

°

C to +40

°

C (-4

°

F to +104

°

F) when drained of all liquid

Humidity ................................................................. 10 % to 90 % non-condensing
Templates .............................................................. Predetermined Test Sequences. Typical capacity 200
Safety ..................................................................... IEC 61010-1: Overvoltage category II, Pollution Degree 2
Electromagnetic Environment ............................. IEC 61326-1: Portable
Emissions Classification...................................... IEC CISPR 1: Group 1, Class A. (Group 1 have intentionally generated

and/or use conductively coupled radio-frequency energy which is
necessary for the internal functioning of the equipment itself. Class A
equipment is suitable for use in non-domestic locations and/or directly
connected to a low-voltage power supply network.)

FCC ......................................................................... CFR47 Part 15.107 and 15.109 Class A
Storage of Results ................................................ Test results stored for later viewing, printing or transfer to PC. Typical

capacity: 250 tests.

Power down ........................................................... The results of tests in progress will be saved in accidental power down
Computer Control ................................................. The Product can be fully controlled using PC HydroGraph and Ansur

software for the Product

Performance Specifications

Flow Rate Measurement

Method ................................................................ Flow is calculated by measuring volume over time
Range ................................................................. 0.1 ml/h to 1500 ml/h (2500 ml/h is shown)
Accuracy ............................................................. 1 % of reading

±

1 LSD for flows of 16 ml/h to 200 ml/h for volumes

over 20 ml, otherwise 2 % of reading

±

1 LSD for volumes over 10 ml

under laboratory conditions. Degassed water at 15

°

C to 30

°

C (59

°

F

to 86

°

F) is recommended for long tests.

Max Test Duration .............................................. 100 hours

Volume Measurement

Method ................................................................ Volume is measured directly by the measuring module in minimum

sample sizes of 60

μ

l

Range ................................................................. 0.06 ml to 9999 ml
Accuracy ............................................................. 1 % of reading

±

1 LSD for flow rates of 16 to 200 ml/h for volumes over

20 ml. otherwise 2 % of reading

±

1 LSD for volumes over 10 ml under

laboratory conditions.

Max Test Duration .............................................. 100 hours

PCA Bolus/Dual Flow Measurement

Method ................................................................ See Volume measurement above
Min Bolus Volume ............................................... 0.5 ml
Resolution ........................................................... 60

μ

l increments

Max Test Duration .............................................. 100 hours

Pressure Measurement

Method (back Pressure and Flow test) ............... Direct measurement of pressure at the inlet port.
Range ................................................................. 0 psi to 45 psi or equivalent in mmHg and kPa
Accuracy ............................................................. 1 % of Full Scale

±

1 LSD under laboratory conditions

Max Test Duration .............................................. 1 hour