Infusion Device Analyzer

Product Connections

7

Product Connections

The Product connects to infusion devices through the front-panel inlet ports. Fluid

drain hoses and accessories connections are made through the rear-panel

connections.

Connect Infusion Devices

As shown in Figure 1, it is recommended that all infusion device connections be

made to the Product front-panel inlet connectors through 3-way stopcocks.

20 ml Syringe
to aid priming

3-way Stopcock

From Infusion
Pumps

gir03.eps

Figure 1. Infusion Device Connections to the Product

The channel 4 inlet shown in Figure 1, shows a 20 ml syringe attached to one

3-way stopcock inlet. The syringe can be used to help priming. It can be used as

shown or can be connected further away from the inlet to help for flow tests. The

syringe can be shared among the channels and removed after the channel is

primed.
Follow these recommendations when you connect to the inlet tubing circuits:

Use adequate prime volumes (for example, 10 ml) to push through any

bubbles.

Use the stopcocks at the inlet to prevent fluid backflow out of inlets between

tests.

When you connect to the inlet circuits (for example, when you attach the

priming syringes to the stopcocks) make sure no new bubbles are

introduced.