Fluke Biomedical IDA-5 User Manual

Page 32

Advertising
background image

IDA-5

Users Manual

22

When ENTER is pushed in the Loading Dose field, the PCA screen changes to

the Prime Mode screen shown in Figure 8. Refer to the Flow Tests section and

prime the channel.

gir18.bmp

Figure 8. PCA Screen - Prime Mode

When the channel is primed, the Prime button is replaced by the AutoStart and

Start buttons. If the basal flow rate is 0.00, select Start to begin the test. Select

Autostart if the basal flow rate is not 0.00.
When the test is in operation, the product monitors basal flow rate, number of

boluses delivered, bolus flow rates, volumes, and durations as shown in

Table 16. Bolus intervals are also monitored. Figure 9 shows a graph of the

supplied boluses and average parameters for all boluses. When a dual flow

pump is tested, it is shown as a single, large bolus followed by the secondary

flow.

Table 16. PCA Screen in Active Mode

gir19.bmp

Menu Control

Description

Graph

Shows a graph of the test. See Figure 9.

Trig

Adds a marker to the saved data when pushed. Push ENTER at the same time as
patient demand on the pump.

End

Stops the test.