Dual flow pump test, Templates – Fluke Biomedical IDA-5 User Manual

Page 33

Advertising
background image

Infusion Device Analyzer

Dual Flow Pump Test

23

Channel 1 PCA/Dual Flow Graph

80

Ave

ml/h

Bolus Count

3

Basal Avg

0

Bolus Avg

65.00

Total Vol

6.00

ml:

15 min

VIEW DETAIL

Press ESC for Status Screen

gir20.eps

Figure 9. PCA Graph Screen

Caution

To avoid inaccurate readings, always repeat a test when

“Bubble” or “Air Lock” is shown on the display while a test is

running. See the “Troubleshooting” section of this manual.

Dual Flow Pump Test

During a test, a dual flow pump can be thought of as a PCA pump that supplies a

large, single bolus.
The primary flow rate must be recorded in the basal flow rate field of the PCA

Information screen. Record the secondary volume in the bolus volume field.

Templates

The Product helps automate and standardize the testing of infusion devices.

Templates are used by the Product to control the sequence of tests done on an

applicable type of infusion device.
Before you can do this, a template for the type of infusion device you will test

must be defined. It is necessary to define a template only once for a particular

type of infusion device. Refer to the device’s manufacturer or service

documentation for necessary test parameters.
The subsequent section shows an example of a generic syringe pump that

requires you test at a flow rate of 20 ml/h for a volume of 10 ml over 30 minutes,

and 50 ml/h for 10 ml, both with a tolerance of 5 %. The occlusion must be tested

at 50 ml/h with an alarm pressure of 12 psi.

Advertising
This manual is related to the following products: