Occlusion Tests

19

When you select Start, the Channel Occlusion screen in Table 13 shows

measurement data as the test continues. Select Graph to show the test data as

a graph.

Table 13. Occlusion Screen in Active Mode

C h a n n e l 1 O c c l u s i o n

Pressure 7.50

psi

Elapsed

Time 00:09:10

Peak Pressure

8.04

psi

Time of Peak

00:08:23

Graph

End

Press ESC for Status Screen

gir14.eps

Measurement Description

Pressure

Shows the amount of pressure and is set by the Operator or the User Defaults.

Elapsed Time

Time since the test was started.

Peak Pressure

Highest pressure sensed since the test started.

Time of Peak

The time that highest pressure was sensed.


The Occlusion Graph screen for the same test is shown in Figure 7. The same

numerical data is shown.

C h a n n e l 1 Oc c l u si o n G r a p h

Inst psi

7.50

Peak psi

8.04

Elapsed Time

00:09:10

Peak

00:07:23

10
psi











0

0 Min 30

View Detail

Press ESC for Status Screen

gir15.eps

Figure 7. Occlusion Graph Screen

Select View Detail to see the Occlusion Detail screen shown in Table 14.