The Product can automatically do tests in a specific manner when specified

conditions are met. Test-stop preferences specifies the default actions that are to

occur. Default values can be changed at the start of each test.
Set which step to edit from the drop-down list in the When field:

Volume

When the set volume is reached. Zero disables

this.

Time

When the set time is reached. 00:00 disables this.

First Condition

Whichever of volume or time is first reached.

Both Conditions

When both the time and volume are reached.

The action when the condition is met is selected from the drop-down list

presented by the Reached then field:

No Action

The test continues. The user is not told.

Notify Operator

A beep sounds and a message shows.

Stop Test Only

The test stops, a beep sounds, and a message

shows.

Stop Test & Save

The test stops and is saved. A message tells you

the test is saved.

The Preferred Parameters let you choose default pressure units and the

position of tolerance lines on flow graphs. Psi, kPa, and mmHg pressure units

are available. The percentage tolerance can be from 0 % to 50 %.

Calibration

The Calibration menu gives access to calibration history and factory functions.

Some items on this menu are available to authorized-service personnel only.

Each screen has instructions on usage. “View Optics” is a diagnostic utility.

Edit Templates

See the “Templates” section earlier in this manual.