Test fluid, Storage, Shipping – Fluke Biomedical IDA-5 User Manual

Page 42

Advertising
background image

IDA-5

Users Manual

32

Test Fluid

The Product is intended to operate with de-ionized water with added detergent.

Fluids intended for use on patients, such as high viscosity, oily, or corrosive

substances will damage the measurement system. Tap water can contain

contaminates which can also damage the transducer.
Test fluid can be made with de-ionized water and a wetting agent such as

MICRO-90. It is recommended that a 0.1 % solution of MICRO-90 in de-ionized

water (preferably degassed) be prepared in volume for daily use and kept in a

sealed vessel. If the water makes too much foam, then a 0.05 % dilution is

recommended.

MICRO-90 is available from:
International Product Corp.

201 Connecticut Dr.

P.O. Box 70

Burlington, NJ 08016-0070 USA

Tel 609 386 8770

International Product Corp.

1 Church Row

Chistlehurst, Kent BR7 5PG United Kingdom

Tel. 0208 467 8944

Storage

Remove all water from the Product before storage, particularly if temperatures

can fall below 5

°

C (41

°

F). Do not pressurize the inlet ports. It is safest to use a

medical suction pump to drain the measuring channels and use the Cycle Valves

from the Calibration menu (follow on-screen instructions).

Shipping

Remove all liquid from the Product before shipping. To prevent liquid from

entering the ports, put the Product in a large plastic bag. Put the bagged Product

into its shipping carton. If this is not available, make sure there is shock

protection with a minimum of 5 cm compressible cushioning inside the carton (for

example, 60 cm x 60 cm x 60 cm).