Flow tests, Average flow, Volume 0.00 – Fluke Biomedical IDA-5 User Manual

Page 24

Advertising
background image

IDA-5

Users Manual

14

Flow Tests

To do a flow test, select Flow on the menu and then ENTER to open the

Channel Flow screen shown in Figure 5. The screen shown is in Prime mode.

C ha nn el 1 F l o w

Average Flow

0.00

ml/h

Volume 0.00

ml

Elapsed Time

00: 00: 00

Inst. Flow

ml/h

Back Pressure

mmHg

Prime

End

Press ESC for Status Screen

gir08.eps

Figure 5. Channel Flow Screen

The vertical bar along the left side of the screen is the Prime indicator of the

liquid in the channel. If there is air in the channel, the indicator will be red. If there

is liquid in the channel, the indicator will be blue. As liquid flows into the channel,

the column changes from red to blue. Red spaces between blue show when the

Product senses bubbles in the measuring system.
When you set up the Product for a flow test, fill the measurement channel with

liquid until the Prime indicator is blue and stays blue. At that time, the Prime

button in the menu at the bottom of the screen changes to AutoStart and a Start

button also appears as shown in Table 8.

Note

The first time a channel is primed after power on, a maximum of
10 ml of liquid can be necessary to fill the measurement channel.
1 ml to 2 ml of liquid can be necessary for subsequent tests on the
same channel. If a 3-way stopcock is used, as discussed in the “How
to Connect Infusion Devices” section, close the inlet of the Product
before you disconnect the infusion device to prevent leaks from the
channel.

Table 8 gives the function of each button in the Flow screen menu.

Advertising