Table 1-3 set point functions, Switch position, Position – Fluke Biomedical 942A-200L-M4 User Manual

Page 10: Jumper

Advertising
background image

942A-200L-M4 & 942A-200L-M5 UDR
Operator Manual

1-6

Table 1-3 Set Point Functions

Switch Position

Setpoint

Units

Factory Setting

0

High Alarm Limit

CPM

1.00E5

1

Warn Alarm Limit

CPM

1.00E3

2

Resolving Time (Tau)

Minutes/count

0.00E0

3

Full Scale Value

CPM

1.00E7

4 Overrange

CPM

1.00E7

5 Conversion

Constant

-

1.00E0

6 Background

Subtract CPM

0.00E0

7

Low Scale Value

CPM

1.00E1

8

Calibration Mode

Seconds

6.0E1

9

Not Used

N/A

E0007 *

A thru E

Not Used

N/A

E0007 *

F Unit

I.D.

N/A

1.00E0

*= Press ENTER push button to clear error code.

Table 1-4 Standard Jumper Placement

Position

Jumper Description

OUT JP1 Reset

2-3

JP2

Standard 27256 PROM

IN

JP3-1

Fixed 1 Second Update

IN

JP3-2

Fixed 1 Second Update

IN

JP3-3

Latches alarms until acknowledged

IN

JP3-4

Gives fail condition without counts, in five minutes

IN

JP3-5

Alarms do not respond to check source counts

1-2

JP4

Negative Input Pulse

1-2

JP5

Negative Shield Polarity

1-2 JP6

Anti-Jam

Active

1-2

JP7

Anti-Jam scintillation type detector