Data bit, Function, Hex data – Fluke Biomedical 942A-200L-M4 User Manual

Page 29: Position

Advertising
background image

Theory of Operation

Main Circuit Board

2

2-15

The following paragraphs describe the operation of each of these registers.

Data Entry (Read Only)

Octal buffer U36 functions as an interface to supply the status of the switches for data entry to the internal
data bus address (4002). The data entry read functions are shown in Table 2-12. The function switch
logic is shown in Table 2-13.

Table 2-12. Data Entry Read Functions

Data Bit

Function

D0

Enter Switch (Active Low)

D1

Value Select Switch (Active Low)

D2

Digit Select Switch (Active Low)

D3 (Not

Used)

D4

D4 – D7 (16 position function switch,

D5

see Table 2-11.)

D6 (Not

Used)

D7 (Not

Used)

Table 2-13. Function Switch Logic

Hex Data

D7

D6

D5

D4

Position

Function

F 1

1

1

1 0

High

E 1

1

1

0 8

Calibrate

Mode

D 1

1

0

1 1

Warn

C 1

1

0

0 9

Undefined

B 1

0

1

1 4

Overrange

A 1

0

1

0 C

Undefined

9 1

0

0

1 5

Calibration

Constant

8 1

0

0

0 D

Undefined

7 0

1

1

1 2

Resolving

Time

6 0

1

1

0 A

Undefined

5

0

1

0

1

3

Full Scale Value

4 0

1

0

0 B

Undefined

3 0

0

1

1 6

Background

Subtract

2 0

0

1

0 E

Undefined

1

0

0

0

1

7

Low Scale Value

0 0

0

0

0 F

Undefined