Operator Manual

2-10

Write Register Decoding

Decoding for write registers within the main circuit board is performed by U2, which is a 1 of 8 decoder.
Control signals for U2 are /REGSELECT, /R/W (active high), /WRITE02 as well as addresses A1, A2, and
A3. U2 decodes two addresses per output, starting at 4000, and ending with 400E. These outputs are
active low. Table 2-4 lists write registers, their assigned function and address.

Address State

A3 A2

A1

Hex
Address

Function

0 0

0

4000

Status

Indicators

0 0

1

4002

E

2

PROM

0 1

0

4004

D/A

Converter

0 1

1

4006

Display

Control

1 0

0

4008

Display

Data

1 0

1

400A

Counter

Control

1 1

0

400C

Relay

Control

1 1

1

400E

(Spare)

Table 2-4. Write Register Functions


The following paragraphs describe the operation of each of these registers.

Counter Control (Write Only)

The counter control register (U43) is an 8-bit register with clear, and responds to address 400A. Upon
initial power-up, the system resets all outputs low. Table 2-5 lists the counter control register write
functions.

NOTE

Data written into U43 remains at the outputs until a reset
occurs or new data is written.

Table 2-5. Counter Control Register Write Functions

Data Bit

Function

D0 COUNTER

CLEAR

D1 COUNTER

ENABLE

D2 HV

SHUTDOWN

D3

HV SELECT (used to digitize HV)

D4

/CLEAR SWITCH LATCH

D5 Not

Used

D6 Not

Used

D7 Not

Used