Jumper, Function, Position – Fluke Biomedical 942A-200L-M4 User Manual

Page 54: Operation, 9 jumper selectable options

Advertising
background image

942A-200L-M4 & 942A-200L-M5 UDR
Operator Manual

3-14

3.9 Jumper Selectable Options

NOTE

The unit must be turned off when changing jumpers.

Table 3-1 Model 942A-200L-M4 and 942A-200L-M5 Option Jumpers (Refer to drawing 942-200-10,

Appendix A, for Jumper Location)

Jumper

Function

Position

Operation

JP1 Microprocessor

Reset

Out

IN

Normal Operation (factory) Momentary
connect to reset microprocessor without
cycling AC power Not Applicable

JP2 PROM

Type

1-2
2-3

PROM 27128
PROM 27256 (factory)

JP3-
1/JP3-2

Statistical Accuracy
(It should be noted that this
accuracy refers only to the
statistical interpretation of
detector counts, and not
detector accuracy)

Out/Out
Out/In
In/Out
In/In

2% Accuracy – 9604
5% Accuracy – 1537 Target Counts
10% Accuracy – 384 Target Counts
Fixed one second display update (factory)

JP3-3 Alarm

Acknowledge

In
Out

Manual Acknowledge (factory)
Automatic

JP3-4

Fail Alarm

In
Out

Enable No Counts Fail Alarm (factory)
Inhibit No Counts Fail Alarm

JP3-5

Check Source Alarm

In
Out

Alarm Inhibited (factory)
Alarm Enabled

JP4

Input Pulse Selection

1-2
2-3

- Input Pulse (Scintillation Detectors)
(factory)
+ Input Pulse (GM Detectors)

JP5

Shield Polarity Selection

1-2
2-3

+ Shield for Negative Pulse (factory)
- Shield for Positive Pulse (GM)

JP6

Anti-Jam Fuse Selection

1-2
2-3
Out

Enable for Normal Operation (factory)
Anti-Jam Circuit fuse bypassed (testing)
Anti-Jam Circuit Disabled (for testing only)

JP7

Detector Type for Anti-Jam
Circuit

1-2
2-3

Scintillation (factory)
GM Tube