Maintenance, calibration and troubleshooting, Adjustment, Purpose – Fluke Biomedical 942A-200L-M4 User Manual

Page 61: 1 maintenance, 2 calibration

Advertising
background image

Maintenance, Calibration and Troubleshooting

Maintenance

4

4-1

Section 4

Maintenance, Calibration and Troubleshooting

4.1 Maintenance

The 942A-200L-M4 or 942A-200L-M5 Universal Digital Ratemeters (UDR) are designed to operate for
extended periods of time with no scheduled maintenance required. If a problem develops, troubleshoot
the UDR per Section 4.3 and the drawings in Appendix B.

4.2 Calibration

Table 4-1 lists the electronic adjustments for the UDR. The procedures for making these adjustments are
described in the following paragraphs. Procedures for calibration of the detector are provided in the
detector instruction manual. Refer to drawing 942-200-10 for component location and drawing 942-200-
13 for the circuit schematic drawing.

Table 4-1. Electronic Adjustments

Adjustment

Purpose

VR1

Zero Adjust/0 – 10 Volt Analog Output

VR2

Gain Adjust/0 – 10 Volt Analog Output

VR3

Gain Adjust V/F Converter (used to set
frequency)

VR4

20 mA Adjust/4 – 20 mA Volt Analog Output
#1

VR5

20 mA Adjust/4 - 20 mA Volt Analog Output
#2

VR6

4 mA Adjust/4 – 20 mA Volt Analog Output #2

VR7

4 mA Adjust/4 – 20 mA Volt Analog Output #1

VR8

Gain Adjust/Signal Input Op Amp

VR9

DC Offset Adjust/Signal Input Op Amp

VR10 High

Discriminator

Adjustment

VR11 Low

Discriminator

Adjustment

VR12

Anti-Jam Threshold Adjust

VR13

Write Cycle Clock Adjust

HV Supply R5

High Voltage Adjust

For calibration of the 0-10 mV and 0-5 Vdc optional analog outputs provided on the 942A-200L-M4 only,
refer to the 942-200-95D instruction manual, provided separately.

0 -10 Volt Output (Factory Set)- Refer to electrical loop drawings S174014D-102,
S174014E1003, S27860A-102 and S321299A-104, Appendix B

1. Disconnect the detector.

2. Set the DVM to the 20 V range.

3. Connect the positive lead of the DVM to P2-10, and the negative lead to P2-11.

4. Adjust VR1 for a DVM reading of 0.00 volts.

Advertising
This manual is related to the following products: