Switch position, 10 udr function switch/function, 11 parameter entry – Fluke Biomedical 942A-200L-M4 User Manual

Page 55

Advertising
background image

Operation

UDR Function Switch/Function

3

3-15

3.10 UDR Function Switch/Function

Table 3-2 UDR Function Switch RDS1 Positions/Function (Refer to drawing 942-200-10, Appendix B,

for Function Switch Location)

Switch Position

Function

0

Selects HIGH Alarm Setpoint

1

Selects WARN Alarm Setpoint

2

Selects Resolving Time (Tau) Limit

3

Selects Analog Full Scale Limit

4 Selects

Overrange

Limit

5 Selects

Calibration

Constant

6 Background

Subtract

7

Selects Analog Low Scale

8 Selects

Calibrate

Mode

9 Not

Used

A-E Not

Used

F Unit

I.D.

3.11 Parameter Entry

Parameter entry is accomplished using a sixteen position rotary switch labeled FUNCTION and three
momentary pushbutton switches labeled ENTER, VALUE, and DIGIT. These switches are located on the
right side of the main circuit board and are accessible by partially removing unit from its mounting case.

The FUNCTION switch is used to select the parameter to be entered. Table 3-3 defines the switch
positions and a description of each parameter can be found in the following paragraphs.

Table 3-3. Function Switch Entry Parameters

Switch Position

Setpoint

Units

Factory Setting

0

High Alarm Limit

CPM

1.00E5

1

Warn Alarm Limit

CPM

1.00E3

2

Resolving Time (Tau)

Minutes/count

0.00E0

3

Full Scale Value

CPM

1.00E7

4 Overrange

CPM

1.00E7

5 Conversion

Constant

-

1.00E0

6 Background

Subtract CPM

0.00E0

7

Low Scale Value

CPM

1.00E1

8

Calibration Mode

Seconds

6.0E1

9

Not Used

N/A

E0007

A thru E

Not Used

N/A

E0007

F Unit

I.D.

N/A

1.00E0

Advertising
This manual is related to the following products: