Appendix

Connector Designations

A

A-5

Table A-9. Standard Jumper Placement

Position

Jumper Description

OUT JP1 Reset

2-3

JP2

Standard 27256 PROM

IN

JP3-1

Fixed 1 Second Update

IN

JP3-2

Fixed 1 Second Update

IN

JP3-3

Latches alarms until acknowledged

IN

JP3-4

Gives fail condition without counts, in five minutes

IN

JP3-5

Alarms do not respond to check source counts

1-2

JP4

Negative Input Pulse

1-2

JP5

Negative Shield Polarity

1-2 JP6

Anti-Jam

Active

1-2

JP7

Anti-Jam scintillation type detector

Table A-10. UDR Jumper Setting Options

Jumper

Connect Description

JP1

-

Momentary connect resets microprocessor

JP2

1-2

PROM type 27128

JP2

2-3

PROM type 27256

JP3-1 IN/OUT

Accuracy

selection

JP3-2 IN/OUT

Accuracy

selection

JP3-3

IN/OUT

Alarm reset control, manual/automatic

JP3-4

IN/OUT

Low count fail control, enable/inhibit

JP3-5

IN/OUT

Alarm when check source is on, inhibit/enable

JP4

1-2

Negative input pulse selection

JP4

2-3

Positive input pulse selection

JP5

1-2

Positive shield (for negative pulse)

JP5

2-3

Negative shield (for positive pulse)

JP6

1-2

Anti-Jam fuse enable – Normal Operation

JP6

2-3

Anti-Jam fuse disable – Test Only

JP7

1-2

Scintillator detector anti-jam input

JP7

2-3

GM tube detector anti-jam