Fluke Biomedical 37-705 User Manual

Page 12

Advertising
background image

Nuclear Associates 37-705
Operators Manual

2-2

2.1.3 Start-up Screen

When the VeriDose V is powered up the startup screen is displayed while the unit performs a self-test.
This test includes a non-destructive, read/write Ram memory test and a RS-232 loop-back test. If an error
occurs, refer to Appendix B for error messages.

2.1.4 Getting Started

This section describes the step-by-step process of making connections to the VeriDose V and calibrating
a diode group for initial measurements. For detailed descriptions of all the VeriDose V features refer to
the MODES section of this manual.

2.1.5 Setup

Connect the power converter supplied with the VeriDose V, 12 VDC @ 1A, to the readout and a
convenient power outlet.

2.1.6 Connecting to a Computer

A PC may be connected to the VeriDose V by connecting the supplied cable, RJ11 (modular phone type
connector) to the RS-232 jack on the VeriDose and the 9-pin female D-Sub to an available com port on
the computer. The interface can easily be tested using either Terminal mode in Windows 3.1,
Hyperterminal in Windows 95 or any suitable communications program. The appropriate com port and the
communications parameters, 19.2k BAUD, 8 data bits, no parity and 1 stop bit are selected via the PC
communication program. Upon power up of the VeriDose V, a startup message will be transmitted to the
computer.

The Excel Add-In supplied with the VeriDose V provides an easy method of downloading the readings
into the Excel spreadsheet. The instructions for installing and running the add-in are found on the
installation disk in the README.TXT file. This file is a text file and may be viewed using a suitable word
processor program.

The Serial (RS-232) port on the VeriDose V may be
used for either a computer connection or the label
printer connection. When the label printer is
selected, via the setup printer menu, computer
communications are suspended. An alternative
printer must be selected to re-enable computer
communications.

2.1.7 Connecting to a Printer - Reference: SETUP-PRINTER

The VeriDose V system produces several types of reports compatible with most printers. The SETUP--
PRINTER menu provides the option of selecting drivers for LaserJet III and later, Label (Seiko Model SLP
220) or ASCII output. The label connects to the VeriDose V RS-232 serial interface located on the right
side of the unit.

2.1.8 Connections

The printer port on the VeriDose V is located on the rear panel and is used for all printer connections
except the label printer. Connect the printer cable to the VeriDose V and the printer.

NOTE