Nuclear Associates 37-705
Operators Manual

2-14

machine used with the selected diode. The COMMENTS field may be used to enter a 7-character alpha-
numeric comment regarding the selected diode. The entries in the database are printed in various reports
for record keeping purposes.

Entering the Database Information:

1. Each channel that has a diode connected must be activated by scrolling to each individual channel

number in the second column and selecting YES in the third column. The fourth column displays
the database items and fifth column allows the user to enter the identifiers for each of these items.
The database items are printed on each report for record keeping purposes.

2. Select channel 1 and enter the Serial number of the diode connected to channel 1. Press the soft-

key below column 5 to scroll the cursor to the right and use the UP and DOWN keys to the right of
the display to enter the appropriate serial number. A blank space can be entered by scrolling down
from "A".

3. Press the OPTION soft-key (column 4) to select the Modality item. Press the soft-key below column

5 and use the UP and DOWN keys to the right of the display to enter the appropriate modality label.

4. Press the OPTION soft-key (column 4) to select the Energy item. Press the soft-key below column

5 to scroll the cursor to the right and use the UP and DOWN keys to the right of the display to enter
the appropriate energy.

5. Press the OPTION soft-key (column 4) to select the Machine item. Press the soft-key below column

5 to scroll the cursor to the right and use the UP and DOWN keys to the right of the display to enter
the appropriate machine label. A blank space can be entered by scrolling down from "A".

6. Press the OPTION soft-key (column 4) to select the Comment item. Press the soft-key below

column 5 to scroll the cursor to the right and use the UP and DOWN keys to the right of the display
to enter the appropriate comment. A blank space can be entered by scrolling down from "A".

7. Repeat this process for each channel that has a diode connected by pressing the column 2 soft-key

to access the related database items.

Calibrate Screen # 7

group
name

NO SER

NUM

AAAAAAA

2

YES

MODALITY ELECTRON

3

ENERGY

000

4

MACHINE

AAAAAAA

5

SAVE CHANGES
TO group name?

GROUP CHANNEL

DIODE

YES

NO


After completing the entry of all the data base information press the ENTER key. Press the YES key
(column 4) to save the database information.