Nuclear Associates 37-705
Operators Manual

2-10

Password

Setup Password screen

MEASURE PRINTER

AAAA

AAAA

SELECT DAT/TIM

SETUP

QAPHANT

CAL

PASSWRD

NAME

DATA

MODE OPTION OLD NEW


This screen permits the operator to change the security password. The operator must enter the existing
password and a new password. The password is a 4 character, alpha-numeric entry. Press the OLD or
NEW soft-key to activate the column, and use the UP and DOWN keys to advance to the appropriate
character. Press the OLD or NEW soft-key to move the cursor to the next character position. Pressing the
ENTER key will save the password entry. The factory default password is AAAA.

Institution Name

Setup Name Screen

MEASURE PRINTER

AAAAAAAAAAAAAAAAAAAAAAA

SELECT DAT/TIM

SETUP

QAPHANT

CAL PASSWRD

NAME

DATA

MODE

OPTION

ENTER INSTITUTION NAME


This screen permits the operator to enter an institution name of up to 23, alpha-numeric characters.
Select the ENTER INSTITUTION NAME soft-key and use the UP and DOWN keys to advance to the
appropriate character. Press the ENTER INSTITUTION NAME soft-key to move the cursor to the next
character position. The institution name is printed on the reports.

Data Format

Setup Data screen

MEASURE PRINTER

US

SELECT DAT/TIM

EURO

SETUP

QAPHANT

CAL PASSWRD

NAME

DATA

MODE OPTION

STYLE


This screen permits the operator to select the data format as US or European. The US format provides
decimal points for the data while the EURO format provides commas in place of the decimal points. Press
the STYLE soft-key scroll to the appropriate format.