Fluke Biomedical 37-705 User Manual

Page 23

Advertising
background image

Operation

Calibrate Modes

2

2-13

Calibrate Screen # 4

group
name

1

NO SER

NUM

AAAAAAA

2

Yes

MODALITY ELECTRON

3

ENERGY

000

4

MACHINE

AAAAAAA

5

SAVE CHANGES

TO group name?

GROUP CHANNEL

DIODE

YES

NO


After completing the entry of all the database information press the ENTER key. Press the YES key
(column 4) to save the database information.

Edit an Existing Group

Calibrate Screen # 5

MEASURE ADD

group
name

SELECT

EDIT

SETUP DELETE

CAL

EXPOSE

REPORT

MODE OPTION

GROUP


The EDIT option may be used to edit the database for the selected group. After selecting the EDIT option,
the operator may scroll through the existing groups by pressing the GROUP soft-key and pressing the UP
or DOWN key pads to the right of the display to scroll through the group names. After the group has been
selected press the ENTER key.

Calibrate Screen #6

group
name

NO

SER NUM

AAAAAAA

2

Yes

MODALITY

ELECTRON

3

ENERGY

000

4

MACHINE

AAAAAAA

5

COMMENT AAAAAAA

GROUP CHANNEL

DIODE

OPTION


After selecting the group to be edited, the items in the database are displayed. The operator may activate
each channel individually, assign a diode to that channel by selecting YES in the diode column, or de-
activate the channel by selecting NO. The database information regarding each diode in the set is dis-
played along the right side of the selected channel. The information is only displayed for active diodes
(indicated by YES in the diode column).

The SER/NUM entry allows the operator to record the serial number of the diode for the selected channel.
The MODALITY (modality) entries, selected by the scroll up and scroll down key, include ELECTRON,
PHOTON and CO60. The ENERGY entry allows the operator to enter the target energy for the selected
diode (the energy at which the diode is calibrated). The MACHINE entry may contain the name of the

Advertising