FLATNESS/SYMMETRY REPORT

VeriDose V, Model number

37-705

Serial number 00000

Firmware revision 1.1 d, March 30, 1998

RADIATION ONCOLOGY

Date: 03/30/1998

Time: 17:36:34

Detector

Group:

GROUP

2

Calibration Date(s):

03/30/1998

03/30/1998 03/30/1998 03/30/1998 03/30/1998

Diode Serial
Numbers:

1932 1933 1934 1935 1936

Machine:

LINAC LINAC LINAC LINAC LINAC

Modality:

electron electron electron electron electron

Energy:

006 MeV

006 MeV

006 MeV

006 MeV

006 MeV

Field Size:

10.0 cm x 10.0 cm

Channel 1

Channel 2

Channel 3

Channel 4

Channel 5

MEASURED
DOSE (cGy)

109.6 109.9 110.2 109.8 110.5

POSITION
ASSIGNMENT

Center -X +X

-Y

+Y

RATIO 1.000

1.003

1.006

1.002

1.008

X-Axis Flatness:

0.274%

Y-Axis Flatness:

0.418%

X-Axis Symmetry:

0.242

Y-Axis

Symmetry: 0.618

Average Measured Dose:

110.0 cGy