Fluke Biomedical 37-705 User Manual

Page 41

Advertising
background image

TREATMENT REPORT

VeriDose V, Model number

37-705

Serial number 00000

Firmware revision 1.1 d, March 30, 1998

RADIATION ONCOLOGY


Patient
Name:

Patient

ID:


Date: 03/30/1998

Time: 17:05:36

Detector

Group:

GROUP

2

Cal Date (s):

03/30/1998

03/30/1998 03/30/1998 03/30/1998 03/30/1998

Diode

S.

N.(s):

1932 1933 1934 1935 1936

Machine:

LINAC LINAC LINAC LINAC LINAC

Modality:

electron electron electron electron electron

Energy:

006 MeV

006 MeV

006 MeV

006 MeV

006 MeV

Elapsed

Time:

59.2

sec

Channel

1

Channel

2

Channel

3

Channel

4

Channel

5

MEASURED
DOSE (cGy)

103.1 93.8 88.8 4.8

3.1

CORRECTION
FACTOR

CORRECTED
DOSE (cGy)

EXPECTED
DOSE (cGy)

PERCENT
DIFFERENCE

ALARM
LEVEL

100.0 100.0 100.0

ALARM LEVEL
EXCEEDED (Y/N)

YES NO NO

DISABLED

DISABLED






Signature:

Advertising