Nuclear Associates 37-705
Operators Manual
B-2
5.
Ch 1
0.0
Ch 4
OFF
104.3
cGy
cGy
Ch 2
0.0
Ch 5
OFF
99.8
cGy
cGy
Ch 3
55.1
ABORT
DONE
If a channel is not zeroed correctly, “ZERO ERR” will be displayed to the right of the appropriate channel.
This could be caused by one of the following:
a. the diode is in a radiation beam while zeroing or
b. the diode may be defective.
Correct the problem and repeat the measurement.