Fluke Biomedical 37-705 User Manual

Page 37

Advertising
background image

ENERGY CONSTANCY REPORT

VeriDose V, Model number

37-705

Serial

number

00000

Firmware revision 1.1 d, March 30, 1998

RADIATION

ONCOLOGY

Date: 03/30/1998

Time: 17:15:33

Detector

Group:

GROUP

2

Calibration Date(s):

03/30/1998

03/30/1998 03/30/1998 03/30/1998 03/30/1998

Diode Serial
Numbers:

1932 1933 1934 1935 1936

Machine:

LINAC LINAC LINAC LINAC LINAC

Modality:

electron electron electron electron electron

Energy:

006 MeV

006 MeV

006 MeV

006 MeV

006 MeV

Normalization/Reference Depth: 1.5 cm

Channel 1

Channel 2

Channel 3

Channel 4

Channel 5

DEPTH
(cm)

1.5 4.3 8.7 11.2 17.4

READING
(cGy)

96.7 112.6 93.9 76.8 38.5

NORMALIZED
(percent)

100.0 116.5 97.1 79.4 39.8

























Signature:

Advertising