ENERGY CONSTANCY REPORT
VeriDose V, Model number
37-705
Serial number 00000
Firmware revision 1.1 d, March 30, 1998
RADIATION ONCOLOGY
Date: 03/30/1998
Time: 17:15:33
Detector
Group:
GROUP
2
Cal Date (s):
03/30/1998
03/30/1998 03/30/1998 03/30/1998 03/30/1998
Diode S.N. (s):
1932
1933
1934
1935
1936
Machine:
LINAC LINAC LINAC LINAC LINAC
Modality:
electron electron electron electron electron
Energy:
006 MeV
006 MeV
006 MeV
006 MeV
006 MeV
Normalization/Reference Depth: 1.5 cm
Channel
1
Channel
2
Channel
3
Channel
4
Channel
5
DEPTH
(cm)
1.5
4.3
8.7 11.2
17.4
READING
(cGy)
96.7
112.6
93.9
76.8
38.5
NORMALIZE
D
(percent)
100.0
116.5
97.1 79.4
39.8