Operation

Calibrate Modes

2

2-15

Delete a Group

Calibrate Screen # 8

MEASURE ADD

group

name

SELECT EDIT

SETUP

DELETE

CAL

EXPOSE

REPORT

MODE OPTION GROUP


The DELETE option allows the operator to delete an existing group. To delete a group from the group
database select the GROUP soft-key and press the scroll up or scroll down key to review the groups.
Once the appropriate group has been selected press the ENTER key.

Calibrate Screen # 9

MEASURE ADD

group

name

SELECT EDIT

SETUP

DELETE

CAL

EXPOSE

REPORT

DELETE group

name

ARE YOU SURE?

MODE OPTION GROUP

YES

No


To delete the selected group, press the YES soft-key, to abort the delete process press the NO soft-key.

Making a Calibration Exposure – Calibrate Screen # 10

MEASURE ADD

group

name

SELECT EDIT

SETUP DELETE

CAL

EXPOSE

REPORT

MODE OPTION GROUP


The EXPOSE option allows the operator to select an existing group and perform a calibration. Select the
GROUP soft-key, scroll to the desired group and press the ENTER key.