4 select mode – Fluke Biomedical 37-705 User Manual

Page 17

Advertising
background image

Operation

Measure Modes

2

2-7

Reports

Report Screen

MEASURE

DOSE TRTFULL

SELECT RATE

FLT/SYM

SETUP ALARMS

E-

CONST

CAL

REPORT

TRTLABL

MODE OPTION TYPE


The report screen is used to select the type of report to be printed. The selected report can be manually
or automatically printed following an exposure. Automatic printing is activated by turning the A-PRINT
function ON (ref. SETUP - PRINTER). To manually print a report based on the last exposure, select
report type and press the ENTER key.

There are four types of reports that can be printed directly from the VeriDose V. The default is the
previously selected report type.

• TRTFULL

Full treatment report

• TRTLABL

Treatment report for label printer only

• FLT/SYM

Flatness and symmetry report

• E-CONST

Energy constancy report


See the Appendices for sample reports.

2.4 Select Mode

Calibration Groups

Select Screen

MEASURE

group name

1 ON

SELECT

2

ON

SETUP

3

OFF

CAL

4 OFF

5

OFF

MODE GROUP CHANNEL

STATUS

MORE


In the SELECT - GROUP mode the operator can select a group of previously calibrated diodes. The UP
and DOWN keys may be used to scroll through the previously entered group names. The MORE soft-key
allows the operator to review the calibration data for each individual diode in the group. The operator
entries for MACHINE, MODALITY, ENERGY, COMMENT, STATUS, SER/NUM, CAL FAC, and DATE
(calibration date) are displayed via the MORE soft-key. When the channel STATUS is being displayed,
the highlighted channel can be toggled ON or OFF by pressing the STATUS soft-key. The highlighted
channel can be changed by pressing the CHANNEL soft-key.