Fluke Biomedical 37-705 User Manual
Page 35
![background image](/files/785451/content/doc035.png)
FLATNESS/SYMMETRY REPORT
VeriDose V, Model number
37-705
Serial number 00000
Firmware revision 1.1 d, March 30, 1998
RADIATION ONCOLOGY
Date: 03/30/1998
Time: 17:36:34
Detector
Group:
GROUP
2
Cal Date (s):
03/30/1998
03/30/1998 03/30/1998 03/30/1998 03/30/1998
Diode S. N. (s):
1932
1933
1934
1935
1936
Machine:
LINAC LINAC LINAC LINAC LINAC
Modality:
electron electron electron electron electron
Energy:
006 MeV
006 MeV
006 MeV
006 MeV
006 MeV
Field Size:
10.0 cm x 10.0 cm
Channel
1
Channel
2
Channel
3
Channel
4
Channel
5
MEASURED
DOSE (cGy)
109.6 109.9 110.2 109.8 110.5
POSITION
ASSIGNMENT
Center -X +X
-Y
+Y
RATIO 1.000
1.003
1.006
1.002
1.008
X-Axis Flatness:
0.274%
Y-Axis Flatness:
0.418%
X-Axis Symmetry:
0.242
Y-Axis
Symmetry: 0.618
Avg. Measured Dose:
110.0 cGy
Signature: