5 setup modes – Fluke Biomedical 37-705 User Manual

Page 18

Advertising
background image

Nuclear Associates 37-705
Operators Manual

2-8

2.5 Setup Modes

Printer

Printer Setup Screen

MEASURE

PRINTER

LASRJET

A-PRINT

ON

SELECT DAT/TIM

ASCII

GRID

SETUP

QAPHANT LABEL TEST

CAL PASSWRD

NAME

DATA

MODE OPTION

TYPE

OPTION

A-
PRINT


The PRINTER setup screen permits selection of the printer type. The LASRJET refers to the Hewlett
Packard LaserJet III and later series printers, and LABEL refers to a Seiko label printer. The ASCII selec-
tion provides a text only output compatible with most printers. The A-PRINT feature, toggled ON or OFF
by pressing the A-PRINT soft-key, allows automatic printing of a report upon completion of an exposure
by pressing the DONE soft-key. The GRID option allows the grid in the report to be turned on or off. This
option applies to ASCII type printer only. The TEST option causes a test report to be sent to the printer to
verify correct printer operation. The test report formats are shown in Appendix A.


When the label printer is selected, the RS-232 port
computer communications (via the RS-232 port)
are automatically disabled. To re-enable computer
communication select the ASCII or LASERJET
printer.

Date/Time

Setup Date/Time Screen

Measure Printer

US

02/25/97

10:15

Select

DAT/TIM

EURO

Setup

QAPHANT

CAL PASSWRD

Name

Data

Mode Option

Style

Date Time


This screen permits the date and time to be entered. The time is entered in 24 hour format. The U.S. style
permits the date to be entered in month/day/year format while the EURO style allows for the day/month/
year format. Press the appropriate column soft-key to move the cursor within the column and the UP and
DOWN keys to increment or decrement the digit at the cursor position. Press the ENTER key to apply the
changes. The time and date setting is used to time/date stamp the measurements and printed on the
reports