6 qa phantom – Fluke Biomedical 37-705 User Manual

Page 19

Advertising
background image

Operation

QA Phantom

2

2-9

2.6 QA Phantom

Setup QA Phantom Screen # 1

MEASURE PRINTER

FLT/SYM

10 cm sq

SELECT DAT/TIM

E-

CONST

SETUP

QAPHANT

CAL PASSWRD

NAME

DATA

MODE OPTION TYPE FIELD


This screen permits data regarding flatness/symmetry reports to be selected. The field size associated
with the measurement can be selected by the operator. The field size selections are 10, 15, 20, 25, and
30 cm sq. When using the VeriDose QC phantom the field size must be set to 20 cm sq. The field size is
printed on the flatness/symmetry report.

Setup QA Phantom Screen # 2

MEASURE PRINTER FLT/SYM

1

xx.x

cm

SELECT DAT/TIM

E-

CONST

2

xx.x cm

SETUP

QAPHANT

3

xx.x

cm

CAL PASSWRD

4

xx.x

cm

NAME

5

xx.x

cm

DATA

MODE OPTION

TYPE

Channel

Depth


This screen permits data regarding energy constancy reports to be entered. The depth associated with
each detector (channel) can be entered by the operator. Select the channel with the CHANNEL soft-key.
Press the DEPTH soft-key to select the first digit of the depth. Repressing the DEPTH soft-key will move
the cursor to the next position. The depth selections can be entered in cm by pressing the UP and DOWN
keypads to the right of the display to scroll through the number set. The depth entry can range from 00.0
cm to 99.9 cm.

The depth entries are printed on the depth dose report. When using the VeriDose QC Phantom enter the
depth for each channel as indicated:

Channel 1

12.5

cm

2

12.5 cm

3

12.5 cm

4 20.5

cm

5

4.5 cm

Advertising