Operation

Calibrate Modes

2

2-17

position. Enter the actual exposure or the sum of multiple exposures (up to 1000 cGy) by pressing the UP
or DOWN keypads to the right of the display to scroll through the number set. After the exposure has
been entered, press the ENTER KEY.

Calibrate Screen # 15

group
name

1 1.157

2

1043

3

1.266

4

5

SAVE CAL

FACTORS?

GROUP CHANNEL

nC/cGy

YES

No


The calibration factor for each channel (diode) is calculated and displayed. "Save calibrations factors?"
appears in the message box. To save the calibration factors, press the YES soft-key, to abort the calibra-
tion process without saving the new calibration factors press the NO soft-key.

Calibrate Screen # 16

MEASURE ADD

SELECT EDIT

SETUP DELETE

CAL

EXPOSE

REPORT

MODE OPTION


The Calibration report is used to print the report. Press the OPTION soft-key and select REPORT. Press
the ENTER key to print the report. Calibration reports for all groups will be printed.