34

5. Operation and Configuration

5.2.4 Adjustable parameters in the program screen

S – Closing time
The closing time controls the duration of the furnace head closing process.

Value range: 00:18–30:00 (mm:ss)

t – Temperature increase rate (for two-stage programs: t

1

)

The temperature increase rate defines by how many degrees per minute the furnace heats up.

Value range °C: 10 – 140 °C/min; Value range °F: 18 – 252 °F/min

T – Holding temperature (for two-stage programs: T

1

)

The holding temperature defines the temperature at which a firing process is conducted.

Value range °C: 100 – 1200 °C; Value range °F: 212 – 2192 °F

H – Holding time (for two-stage programs: H

1

)

The holding time indicates how long an object is fired at the holding temperature.

Value range: 00:00 – 60:00 (mm:ss)

Vacuum on (for two-stage programs: V1

1

)

The parameter defines the temperature at which the vacuum is activated.

Value range °C: OFF or 1 – 1200 °C; Value range °F: 0 or 34 – 2192 °F

Vacuum off (for two-stage programs: V2

1

)

The parameter defines the temperature at which the vacuum is deactivated.
Holding time without vacuum: If V2 is set one degree lower than the holding temperature, the vacuum is

ended before the holding time.

Holding time with vacuum: If V2 corresponds to the holding temperature, the vacuum is maintained

during the entire holding time.

Long-term cooling with vacuum: If V2 is set one degree higher than the holding temperature, the

vacuum is maintained during long-term cooling.

Value range °C: OFF or 1– 1200 °C; Value range °F: 0 or 34 – 2192 °F

Long-term cooling
If long-term cooling is activated, the furnace cools to the set temperature (L) at the end of the holding
time with the furnace head closed.

Value range °C: OFF or 50 – 1200 °C; Value range °F: 0 or 122 – 2192 °F

Cooling rate
Can only be set if long-term cooling "L" is activated. The cooling rate defines by how many degrees per
minute the furnace cools down.

Value range °C: Off or 1 – 50 °C/min; Value range °F: Off or 2 –90 °F

t2 – Temperature increase rate Stage 2
This parameter defines for the second temperature level by how many degrees per minute the furnace
heats up.

Value range °C: 10 – 140 °C/min; Value range °F: 18 – 252 °F/min

T2 – Holding temperature Stage 2
The holding temperature for the second temperature level defines the temperature at which a firing
process is conducted.

Value range °C: 100 – 1200 °C; Value range °F: 212 – 2192 °F

H2 – Holding time Stage 2
The holding time for the second temperature level indicates how long an object is fired at the holding
temperature.

Value range: 00:00 – 60:00 (mm:ss)

Vacuum on Stage 2
The parameter defines the temperature at which the vacuum for the second temperature level is activated.

Value range °C: Off or 1 – 1200 °C; Value range °F: 0 or 34 – 2192 °F

Vacuum off Stage 2
The parameter defines the temperature at which the vacuum for the second temperature level is
deactivated. If V2

2

corresponds to the holding temperature, the vacuum is maintained during the entire

holding time.

Value range °C: Off or 1 – 1200 °C; Value range °F: 0 or 34 – 2192 °F