Furnace head cannot be
initialized

The furnace head cannot be brought to the end position. It might be
blocked by external mechanical influences! If this is not the case, contact
your local Ivoclar Vivadent Service Centre.

28
**

The furnace head does not reach
the target position

Furnace head does not open/close correctly. The furnace head is obstructed
or has been moved by hand. Operate the furnace head only with the keys
intended for that purpose.

32
**

no

The vacuum is not released

The vaccum cannot be released. The vacuum valve might be stuck or dirty.
Contact a Service Technician.

33

Necessary vacuum (xxxmbar) is
not reached within 1 min

The vacuum cannot be built-up. Check the seal of the firing chamber,
vacuum hose, vacuum pump, pump fuse.

106

Timer active – device is in standby The program cannot be started, because the unit was put into stand-by

mode by the timer (heating turned off). Deactivate the timer or extend the
time frame to be able to execute programs.

109

Incorrect value for HV or V2

Enter plausible values for HV or V2.

110

HV > H (H2)

Enter a lower value for HV or a higher value for H (H2).

111

Max. number firing protocols
reached

The maximum number of firing program protocol entries has been reached.
A repeated protocol will delete / overwrite existing entries.

150

Memory error

Internal memory error. Please restart the device.

531

yes

Error while trying to store burn-
data

An error occurred while logging firing program data. The storage medium
may be full.

702

Brief power failure during a pro-
gram in progress

A program in progress was interrupted by a short power outage. The
program is being continued.

800

Final vacuum value not reached

The set vacuum end value could not be achieved. Check the vacuum
pump.

801

Vacuum drop

An unacceptable vacuum drop has occurred.

802

The vacuum does not increase
(self-test)

A vacuum increase could not be measured! Check the following points: Is
the firing chamber tight (no contamination on the sealing surfaces)? Is the
vacuum hose connected? Is the vacuum pump connected? Is the fuse F1 in
order?

803

The vacuum system is not tight

The tightness of the vacuum system is not given. Check the sealing
surfaces for possible contamination.

1302
**

ATK2-calibration:
preheat 962°C

Error during calibration. Sample might be incorrectly inserted. Try again
with a new sample and make sure the sample shows the correct contact.

1501

yes

Hint -> heating muffle

The heating muffle has ****working hours. Please perform a heating test
(diagnosis) and follow the instructions on the display and in the Operating
Instructions.

1510

Temperature > VT at the start of a
firing program

The temperature of the firing chamber is higher than the predrying
temperature. To continue the programm nevertheless, press the Start
button.

1522

SW-update:

An error occurred during the software update. Do NOT switch off the
furnace. Try again. If the error reoccurs, try an update via the USB interface.

1541

Error during copying a file

An error occurred while copying the file. Possible causes: File too large,
invalid data type

1800

yes

USB stick

No USB memory stick available.

1815

yes

USB stick

There is no USB stick available or the stick is not prepared as an external
program memory. The USB stick can be prepared in the settings menu.

1820

yes

USB stick

The USB memory stick is already prepared for programs.

1825

yes

USB stick

Please unplug all USB devices and then just plug in the USB memory stick
to be prepared.

1830

yes

USB stick

Not enough free memory.

1835

yes

Media playback

Playback of the file is not possible.

1900

yes

Data backup – recover

The new software contains setting options which were not yet available at
the time of the data backup. Only the existing settings of the data backup
were restored.

1901

yes

Data backup – recover

The new software contains program parameters which were not yet
available at the time of the data backup. Only the existing program
parameters of the data backup were restored.

8. What If …