Product specifications – Ivoclar Vivadent P710 User Manual

Page 83

Advertising
background image

83

9. Product Specifications

9.1 Delivery form

– Programat P710
– Power cord
– Vacuum hose
– Programat Firing Tray Set 2
– Automatic Temperature Checking Set ATK 2 (test set)
– USB download cable
– Programat WLAN Kit
– Programat Bluetooth Kit
– DSA Holder Kit
– var. accessories

Recommended accessories:
– Temperature Checking Set ATK 2
– Vacuum pump VP5, VP3 easy

9.2 Technical data

Power supply

110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz

Overvoltage category

II

Contamination level

2

Acceptable voltage fluctuations

± 10 %

Max. power consumption

14 A at 100 V
12 A at 110 – 120 V
8.5 A at 200 – 240 V

Acceptable data for pumps of other manufacturers:
Max. output
Final vacuum

250 W /max. leakage current 75 mA
< 50 mbar
Only tested pumps should be used.

Electrical fuses

110 – 120 V:
125 V /T15 A (heating circuit)
125 V / T 5 A (vacuum pump)

200 – 240 V:
250 V / T8 A (heating circuit)
250 V / T 3.15 A (vacuum pump)

Dimensions of electrical fuses

110 – 120 V:

Diameter 6.3 x 32 mm

200 – 240 V:

Diameter 5 x 20 mm

Dimensions of the closed furnace

Depth: 495 mm
Width: 320 mm / 395 mm (with cooling tray)
Height: 320 mm

Usable size of the firing chamber

Diameter: 90 mm
Height: 80 mm

Max. firing temperature

1200 °C (2192 °F)

Weight

18.7 kg

Advertising