Operation and configuration – Ivoclar Vivadent P710 User Manual

Page 47

Advertising
background image

47

User code
The user code can be individually changed.

It is recommended to make a note of the changed user code and keep it separately. If forgotten,

the user code may only be reset with the help of Ivoclar Vivadent After Sales Service.

Setting option: 1000 to 9999
Operating mode
Setting of the desired operating mode. See Chapter 5.4.8 Operating mode for details.

Setting option: Standard / Safe / Production
Furnace number
Here, you can enter any furnace number. If the operating mode "Production" is activated, the number will
be prominently displayed in the screen.

Setting option: 1 to 99
Protocolling
If this function is activated, the program data are saved in a protocol entry after every firing procedure.
The following protocol settings are available:

Inactive: Protocolling is not active

Printer: At the end of the program, the parameters used are logged and saved in the furnace. Additionally,
the protocols are printed on a connected USB printer.

PC: At the end of the program, the parameters used are logged and saved in the furnace. If the furnace is
connected to the PrograBase Software, the saved table entries are synchronized with the connected
laptop/PC. Protocols can be edited, saved and printed by means of the PrograBase software.

Table: At the end of the program, the parameters used are logged and saved in the furnace. The data can
be viewed via the Diagnosis menu.

Setting option: Inactive / Printer / PC / Table
Laboratory name
The laboratory name can be entered here. It is automatically added to the protocols.

Setting option: Laboratory name entry
Calibration interval
Setting for the notification as to when the next calibration should be conducted.

Setting option: 1 / 3 / 6 / 12 months
Set heating muffle firing hours to zero
If this function is executed, the heating muffle firing hours are set to "zero". This function can only be
executed by entering the user code.

Setting option: execute
Set vacuum pump hours to zero
If this function is executed, the vacuum pump hours are set to "zero". This function can only be executed
by entering the user code.

Setting option: Execute
Reset to factory settings
If this function is executed, all programs and settings are reset to the status before the initial start-up. This
function can only be executed by entering the user code.

Setting option: Execute
Preparing USB stick programs
If this function is executed, a USB stick is prepared as program memory.
Loading and individual start screen
This function is used to load an individual start screen from a USB stick. Once an individual start screen has
been loaded, it will be shown for a few seconds when the furnace is switched on the next time.
Loading individual tune
This function is used to load an individual tune. This melody is played as a speaker sound as described in
Chapter 5.1.6.