TE Technology TC-36-25-RS485 User Manual

Page 15

Advertising
background image

 

15

 

Thermistor attached to cold plate of cooler. 

 

The standard thermistor supplied with the TC‐36‐25 RS485 is the MP‐3193.  Appendix C “Thermistor Styles for TC‐
36‐25
” shows dimensional data on the MP‐3193 and other thermistor styles readily available as well as the 
corresponding temperature‐resistance data (see also 

www.tetech.com

 for other thermistors that may be 

available).  If you are using the MP‐3193, it is recommended that you use thermal grease (such as TE Technology 
TP‐1) between the thermistor and the surface to which it is mounted. 
 
See also section 2.3 for further details on using thermistors with different resistance‐temperature curves than the 
standard thermistor. 

 
1.3

Connect the thermistor wire leads to terminals JP2‐5 and JP2‐6. The thermistor is non‐polarized, so it does not 
matter which particular lead goes to which terminal. (See “Controller Wiring Diagram” for reference. The 
thermistor is indicated in the diagram as the ‘primary thermistor’.)  You can lengthen the wire leads if necessary.  
The gauge of the wire is generally not important since the resistance of the wire is insignificant to the resistance of 
the thermistor. 

 
1.4.0 The TC‐36‐25‐RS485 can be used with either one or two separate power supplies, depending primarily on the 

nominal operating voltage of the TE device. If the maximum TE device input voltage is less than 12 V, then two 
power supplies must be used with the controller. In this case, one power supply is for the TE device, and the other 
power supply is for the controller itself (the microprocessor and associated electronics). Of course, just one power 
supply can be used for powering both the controller and the TE device provided that the TE device’s nominal 
operating voltage is within (12 to 36) V. 

 
 

When using one power supply for powering the controller and the TE device together, the power supply 
input voltage is passed directly through the controller to the TE device during the "ON" pulse.  The user 
should choose an input voltage that is required for the TE device and yet is also ≥12 V but ≤36 V. The 
controller could be damaged if it is operated outside this range. 

When using two power supplies, the controller input power supply voltage must be ≥12 V but ≤36 V and 
provide a minimum of 200 mA current. The TE device power supply input voltage can be ≥0 V but ≤36 V, 
but it should be no greater than the rated input voltage for the TE device. 

The TE device current rating should be no more than 25 A at the supplied input voltage regardless of 
whether you are using one power supply or using two independent power supplies. NOTE:  the maximum 
allowable ambient temperature for the controller is 60 °C.  The maximum allowable temperature of the 
controller base (underneath the transistor‐mounting area) is 90 °C. The controller might need additional 
heat sinking, depending on ambient conditions and how much current is being drawn by the TE device.
 