TE Technology TC-36-25-RS485 User Manual

Page 46

Advertising
background image

 

46

6. If the temperature controller receives the command and the checksum is correct, it will send back:  

(stx)000000fae7(ack).  The e7 at the end of the string is the checksum of “000000fa”.  The value 
000000fa

16

 converts to 250

10

.  This number must then be divided by 100

10

, so the value would be 

interpreted as 2.50 °C.  As in the other examples, if the checksum from the query were not correct the 
temperature controller would send back:  (stx)XXXXXXXXc0(ack).   

 
III.  Communications Parameters (JP4 connector on controller board) 
 

 

 

 

RS485 

 

 

 

 

Baud Rate ‐ 115200 

 

 

 

 

Data Bits ‐ 8 

 

 

 

 

Parity ‐ None 

 

 

 

 

Stop Bits ‐ 1 

 

 

 

 

Flow Control ‐ None 

         
IV.  Commands 
 

1. INPUT1 (reads the temperature of the primary thermistor) 

Write Command: NA 
Read Command:  01 
Interpret: 

Convert the hexadecimal “return value” to decimal and divide by 100

10

 to interpret as 

the temperature in °F/°C. 
Example: 
Suppose the temperature is 10.00 °F (sensor type = 1), and the address is 62

16

Send: *62010000000049^ 
Receive: *000003e8c0^ 
Convert 000003e8

16

 to decimal = 1000

10 

Divide by 100 = (10.00 °F)  

 

2. DESIRED CONTROL VALUE (set value) 

Write Command: NA 
Read Command:  03 
Interpret:  

This command returns the set value determined by Input2 or as a fixed value set by 
communications. Convert the hexadecimal “return value” to decimal and divide by 
100

10

 to interpret. 

 

3. POWER OUTPUT 

Write Command: NA 
Read Command:  02 or 04  
Interpret: 

‐511

10

 represents –100% output. 

0 returned is 0% output. 
511

10

 represents 100% output. 

 