TE Technology TC-36-25-RS485 User Manual

Page 45

Advertising
background image

 

45

7. If the temperature controller receives the command and the checksum is correct, it will send back:  

(stx)000003e8c0(ack).  If the checksum is not correct the temperature controller will send back:  
(stx)XXXXXXXXc0(ack).   

 
C) Send the set temperature of ‐1.50 °C. 

1. The controller address, AA, is in this example 62

16

2. The control command, CC, for “FIXED DESIRED CONTROL SETTING” is 1c. 
3. Multiply the desired set‐point temperature by 100

10

 (‐1.50 x 100 = ‐150). 

4. Convert ‐150

10

 to hexadecimal by taking the two’s complement of the number. (This can be easily done 

by using the formula 2

n

 – N, where n is the word‐bit size and N is the absolute value of the number 

being converted.) Hence, you have 2

32

 ‐ 150 = 429496714610. This value then converts to ffffff6a

16

5. Compute the checksum (SS) by adding the hexadecimal ASCII values of the following characters:  6, 2, 1, 

c, f, f, f, f, f, f, 6, and a: 

          

 

ASCII Character:   

 

 

Hexadecimal Value: 

 

 

 

 

 

 

 

36 

 

 

 

 

 

 

 

32 

 

 

 

 

 

 

 

31  

 

 

 

 

 

 

 

63  

 

 

 

 

 

 

 

66  

 

 

 

 

 

 

 

66  

 

 

 

 

 

 

 

66  

 

 

 

 

 

 

 

66  

 

 

 

 

 

 

 

66  

 

 

 

 

 

 

 

66  

 

 

 

 

 

 

 

36  

 

 

 

 

 

 

 

61  

 

 

 

Sum 

 

 

 

 

3f7

16

  

 

 

 

The 8‐bit checksum is the 8 least significant binary bits of the sum, represented as f7 in hexadecimal. 

6. Combining all of these characters in one string we send:  (stx)621cffffff6af7(etx). 
7. If the temperature controller receives the command and the checksum is correct, it will send back:  

(stx)ffffff6afb(ack).  If the checksum is not correct the temperature controller will send back:  
(stx)XXXXXXXXc0(ack). 

 
D) Read the actual temperature of the control thermistor (INPUT1) from the controller at address 62

16

1. Let us assume the actual temperature is 2.50 °C, and the working units have been defined as °C.  This 

means that the controller will return a value in °C. 

2. The controller address, AA, is in this example 00. 
3. The control command, CC, for “INPUT1” is 01. 
4. There is no send value, so we can just calculate the checksum (SS) by adding the ASCII values of the 

following characters:  6, 2,0,1,0,0,0,0,0,0,0,0: 

         ASCII Character:   

 

 

Hexadecimal Value: 

 

 

 

 

 

 

 

36  

 

 

 

 

 

 

 

32 

 

 

 

 

 

 

 

30 

 

 

 

 

 

 

 

31 

 

 

 

 

 

 

 

30 

 

 

 

 

 

 

 

30 

  

 

 

 

 

 

 

30 

 

 

 

 

 

 

 

30 

 

 

 

 

 

 

 

30 

 

 

 

 

 

 

 

30 

 

 

 

 

 

 

 

30  

          

 

 

 

 

 

 

30  

Sum 

 

 

 

 

0249

16

  

The 8‐bit checksum is the 8 least significant binary bits of the sum, represented as 49 in hex. 

5. Combining all of these characters in one string we send:  (stx)62010000000049(etx). 