Caution – Amprobe DM-4 Power-Quality-Recorder User Manual

Page 37

Advertising
background image

3) Turn ON the unit by pressing the ON/OFF push button

4) If you want to measure current and power, connect the clamp meter to the phase

conductor respecting the specifications shown on the clamp and the connections

shown in the picture.

5) Apply voltage to the electrical equipment under test (if previously shut off for the

instrument connection).

6)The values of the available electrical parameters will be displayed on the Remote

User Interface (PDA) display or the PC screen. For further details see the Remote

User Interface (PDA) settings.

7) If you want to record, simply press the RECORD push button on the device.

11.2 USING THE INSTRUMENT IN A THREE PHASE 4- WIRE SYSTEM (WYE)

m

Instrument connection in a three-phase 4 wire system

36

DM-4 Power Quality Recorder

CAUTION

The maximum voltage between L1, L2, L3 & COM inputs is CAT III 635

V~ phase – phase 370 V~ phase to earth. Do not measure voltages

exceeding the limits prescribed by this manual. Should you exceed the

voltage limits you could damage the instrument and/or its components

or endanger your safety.

Phase 1

Phase 2

Neautral

Phase 3

Black

Red

Yellow

Blue