Caution – Amprobe DM-4 Power-Quality-Recorder User Manual

Page 38

Advertising
background image

m

1) Check the cables connection.

2) Connect the phase and neutral voltage wires respecting the connections shown

in the picture.

3) If you want to measure current and power, connect the clamp meter to the phase

conductor respecting the specifications shown on the clamp and the connections

shown in the picture. In case of doubts select the position POWER and,

connecting one clamp at a time, check if:

a) The phase sequence is correct

b) The active power P of each phase is positive.

c) The value of the Pf of each phase is not excessively low (typically it’s not lower

than 0.4). In case the Pf is lower than 0.4, check if the phase voltage is

associated to the right clamp meter (for example the voltage of phase 1 must

be associated to the clamp meter no. 1).

4) Apply voltage to the electrical equipment under test (if previously shut off for the

instrument connection).

5) The values of the available electrical parameters will be displayed.

6) If you want to record:

a) Check and, if needed, modify the values of the basic parameters (see

paragraphs 5.2 and 5.3).

b) Check and, if needed, modify the recording parameters.

c) To start the recording press RECORD.

37

DM-4 Power Quality Recorder

CAUTION

If possible, before connecting the instrument to the electrical equipment

to be tested take the power supply off the electrical equipment.

Advertising