Amprobe AMB-110 Insulation-Resistance-Tester User Manual

Page 13

Advertising
background image

8

To clear all memory locations:
1. Select Configuration from the main menu
2. Highlight the Memory Clear option using and arrows.
3. Press the SELECT key, (“Press MEM to confirm!” message will be displayed).
4. Press the MEM key to clear all memory locations or ESC to cancel the activity.

Fig. 4 Configuration state

Parameter

Value

Note

Contrast

0%..100%

Adjustment of the LCD contrast

Filter

Fil1, Fil2, Fil3, Fil0

Selection of noise rejecting filter, see the description of
Filter Option

Time

Set real time (hour: minute)

Date

Set current date (day-month-year)

Com Port

RS 232 2400,RS 232 4800,
RS 232 9600,RS 232 19200,
USB 115000

Set communication mode and rate.

Memory clear

Clear all memory locations

Initialization

For internal factory and service maintenance only!

DIAG. Starting time

0%..90%

Adjustment of start of the timer in the DIAGNOSTIC
TEST functions, according to the nominal voltage
Unominal. See additional explanation in Diagnostic
test.

Table 1. Configuration parameters

MEASUREMENTS
Generally Information about High DC voltage testing

See page 4 WORKING WITH THE INSTRUMENT for safety warnings.

The purpose of insulation tests
Insulating materials are important parts of almost every electrical product. The material’s properties
depend not only on its compound characteristics but also on temperature, pollution, moisture,
ageing, electrical and mechanical stress, etc. Safety and operational reliability require the regular
maintenance and testing of the insulation material to ensure it is kept in good operational
condition. High voltage tests are used to test insulating materials.

Advertising
This manual is related to the following products: