The Fig. 10 shows states when Graph R(t) is enabled. After taking a measurement, you can switch
back and forth between numerical mode display and graphical mode display by pressing or keys.

graphical mode

numerical mode

Note

• It is not possible to switching mode of presentation when measurement running!!!

Initial display -

numerical mode

Initial display -

graphical mode

Display with results -

numerical mode

Display with results -

graphical mode

Fig. 10 Insulation Resistance function display states - Graph R(t) enabled

Measurement procedure:

• Connect the test leads to the instrument and to the test object.
• Select INSULATION RESISTANCE function in MAIN MENU.
• Press the START/STOP key and release it, continuous measurement will begin.
• Wait until the test result has stabilized then press the START/STOP key again to stop the

measurement or, if enabled, wait for the set timer runs out.

• Wait for the object under test to discharge.
• The result can optionally be saved pressing the MEM key twice, see the description of Store,

Recall and Clear Operation.