Amprobe AMB-110 Insulation-Resistance-Tester User Manual

Page 27

Advertising
background image

22

DD on/off

ON

ON: DD enabled, OFF: DD disabled

Graph R(t)

ON

Enable/Disable Graph R(t)

Time1, Time2 and Time3 are timers with the same start point. The value of each presents the
duration from the start of the measurement. The maximum time is 99 min. The following Fig.17
shows the timer relationships.

0

Time1

Time2

Time3

t

Start

R15s

R01min

DAR

R03min

PI

Stop

(if enabled DD)

Time1 Time2

Time2 Time3


Enable/Disable the graph R(t) and Set-up the graph R(t) parameters in the Diagnostic Test function:

• Press the SELECT key, Set-up menu appears on display, see the Fig. 18.
• Select the parameter Graph R(t) to be set using the and keys;
• Enable/Disable the graph R(t) using the and keys.
• Press the SELECT key to Set-up the parameters of the graph R(t), see the Fig. 19. Press the ESC

key to return to basic Set-up menu in the Diagnostic Test function.

• Complete the set-up adjustments pressing either the ESC key or START/STOP key (to run the

measurement directly). The last displayed settings are stored.

Fig. 18 Set-up menu in

Diagnostic Test measurement

Fig. 19 Set-up menu of the

Graph R(t)

Note

• The time duration of Graph R(t) is equal to the value of Timer 3.
• The Timer value could be very long (up to 99 minutes), so the Special automatic decimation

algorithm is use to write the Graph R(t) to the LCD.

• The cursors of the Graph R(t) could be activated with key
• The cursors of the Graph R(t) could be moved with and keys.

Warning!

• Refer to Warnings chapter for safety precautions!

Advertising
This manual is related to the following products: