Adjustmentandcommissioning

Run the teach-in:

Open the valve to fill the tank.

Save the new K
factor calculated or
not.

Vol.Teach

Sensor

Start Teach

Calib

K teached:

K=3,810

Cal. Result

BACK

Save: Yes/No

Cal. Result

The device displays the K factor

calculated by teach-in.

OK

The device displays the instantaneous
flow of the fluid.

OK

K teached set to

max K=9999

BACK

Confirm the K factor =9999 with “YES”
or keep the current K factor with “NO”.

If the calculated K factor is

≥ 9999, it is forced to 9999.

Meas. Progr.

When the tank is filled, close the valve.

OK

2.001 l/s

The device displays the volume calcu-

lated with the current K factor. If the

display shows “++++”, the volume

calculated > 9999 volume units: conti-

nue the procedure as normal.

Enter the actual volume of fluid
in the tank.

Input Vol.

101.2 l
+099.0 l

Save modified

data? Yes/No