65

Maintenanceandtroubleshooting

red
leD

current
output

transistor
output

icon message dis-

played in the
info menu

possible cause

recommended action

OFF

4-20
mA

Switched

2)

"W:TRnPu 1:1
set"

In “Pulse” mode, the
volume entered for
a pulse is incorrect
(see chap. 9.12.10).

Enter an appropriate
volume.

check the K factor.

1)

if the DIAG MODE function of the "Output.AC1" or "Output.AC2" menu is set to "22 mA" (see chap. 9.12.9); else, the

current output delivers a standard current between 4 and 20 mA

2)

If the "PVAR" function of the "Output.TR1" and/or "Output.TR2" menus is set to "warning" (see chap. 9.12.10); else, the

transistor outputs are operating depending on the set thresholds.

English

Type 8026- 8036- 8076