Checking the outputs behaviour, Information menu – Burkert Type 8076 User Manual

Page 64

Advertising
background image

62

Adjustmentandcommissioning

9.15.3. checking the outputs behaviour

See chap. 9.10 to access the Test menu.

The feature allows for simulating the measurement of the physical parameter or the value of the totalizer to check
if the outputs are correctly configured.

The

T

icon is displayed in place of the

icon as soon as the check for the correct working of an output

has started. During the check the related output does not react according to the measured physical value.

PVar:

Value:

INPUT

Flow_L

Flow_m3

Flow_gal

Flow_Igal

Freq.

Tot1_L

Tot1_m3

Tot1_gal

Tot1_Igal

Tot2_L

Tot2_m3

Tot2_gal

Tot2_Igal

Test

Sensor

PVAR: choose the physical parameter or the totalizer to be tested.
VALUE: enter a physical parameter value by modifying the flow time unit or a totalizer value, depending on the
choice made in the “PVAR” function above, to check output behaviour.

9.16. information menu

See chap. 9.10 to access the Info menu.

Error

Info

Warning

Smiley

Software

Main

Sensor

Soft.Main

E:MESSAGE

W:MESSAGE

MESSAGE

Not set

This menu makes it possible to read:

• a short description of the cause that generated an event related to the following icons, when they are displayed

by the device:

- ERROR:

ERR

- WARNING:

- SMILEY:

or

See also "If you encounter problems" at chap. 10.3

• the software version (function “SOFTWARE”) of the acquisition/conversion board for the physical parameters

measured.

English

Type 8026- 8036- 8076