78

Operatingandfunctions

9.7.3

entering the K factor of the fitting used

The device determines the quantity of fluid that flows through the pipe using the fitting K factor.

The K factor of the fitting used can be entered here. The device may also determine the K factor using a teach-in
procedure: see chap. “9.7.4”.

The device uses the new K factor as soon as "SAVE YES" is confirmed when leaving the Parameters
menu.

The K factor of the fitting used can be found within the operating instructions of the fitting.

The operating instructions of the Bürkert fittings can be found on the CD delivered with the device or on
the internet at

www.burkert.com

.

The display shows the K factor
of the fitting, last entered or
determined using a Teach-In
procedure.

Confirm.

K-FACtOR

K=10.00

REtURN

Enter the K factor (value between 0,01 and
99999.9) of the fitting used.

Confirm the displayed value.

K=2.85

OPtiONAL

VALiD N/Y

Is only displayed if the K factor has been changed.

Confirm or not the entered K factor.

Figure 57:

Entering the K factor of the fitting used

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

9.7.4

Determining the fitting K factor using a teach-in procedure

The device determines the flow rate in the pipe using the fitting K factor.

The "TEACH V." or "TEACH F." parameter allows the device to determine the fitting K factor using a Teach-In
procedure. The K factor may also be directly entered: see chap. “9.7.3”.

The teach-in can be done either depending on a known volume ("TEACH V.") or depending on the flow rate
("TEACH F.") in the pipe, measured by a reference instrument.
Determining the fitting k factor using a teach-in procedure depending on a volume ("teAch v.")

• Before starting the Teach-In procedure, connect a valve to the relay output DO2.

• The device uses the new K factor as soon as "SAVE YES" is confirmed when leaving the Parameters

menu.

English

Type 8025 - 8035 BATCH