Density measurement, Accuracy – Fluid Components International CMB User Manual

Page 16

Advertising
background image

3.4.3 Density

measurement


The attainable accuracy depends on the selected calibration type.

Without calibration no density measurement is possible and the empty pipe recognition
is not available!

Model

without

3-Point

5-Point

CMB-B

CMB-C
CMB-D

5 g/l

3 g/l

CMB-E

5 g/l

3 g/l

CMB-F

5 g/l

3 g/l

CMB-G

5 g/l

3 g/l

CMB-J

5 g/l

3 g/l

CMB-K

5 g/l

3 g/l

Density accuracy

not available

no m

easurem

ent of d

ensity

3.4.4 Accuracy

Mass flow

Accuracy CMB-B to CMB-K

± 0.15% of actual flow + zero point stability (see Section 3.4.2 CMB
flow ranges)

Repeatability error

± 0.05% of actual flow (sensor with transmitter)

Additional measured values

Volume flow

± 0.2 % of actual value + zero point stability

Temperature

± 0.5 °C

Hysteresis n/a
Settling time

1 to 15 seconds

Startup drift

15 minutes

Long-term drift

± 0.02 % of upper-range value per year

Influence of ambient temperature

± 0.005 % per K

Influence of fluid temperature

Compensated

Influence of fluid pressure

For fluids: too small to be relevant


Page 16 of 106

CMB & CT OPERATING MANUAL