10. CT operating conditions

10.1 Installation conditions and cable glands

The integral mount version of the CT transmitter in the SG1 housing is to be installed in accordance with
Section 3.5.1 Installation on page 18. If the CT transmitter is installed separately, a vibration-free installa-
tion site must be ensured.

Warning:
Additional cable glands:
They are not contained in the scope of supply. The operator is responsible for
the fact that according to the enclosure and ignition enclosure certified cable
glands or screws are used. The kind of the thread is stamped on the rating
plate.
At the connection between sensor and transmitter a metalized cable gland
must be used for the screen.
(See 11.5.2.2 “Wiring diagram for the remote mount configuration of sensor
and CT” page 42)

10.2 Environmental conditions

10.2.1 Ambient

temperature

− 20 °C to + 60 °C (-4 °F to 140 °F), below 0 °C (32 °F) the readability of the LC display will be limited.

10.2.2 Ambient temperature range

− 20 °C to + 60 °C (-4 °F to 140 °F)

10.2.3 Storage

temperature

−25 °C to + 60 °C (-13 °F to 140 °F)

10.2.4 Ingress

protection

Standard housing SG1, IP 68 (NEMA 6P)
Explosion-proof electronics housing
Terminal compartment: with terminals and “Increased safety” type of protection.

Warning:
Ingress protection IP 68 is only achieved if suitable and tightly screwed down
cable glands or conduit are used. If the cable glands are only tightened manu-
ally water may leak into the terminal compartment in the housing.

Danger:
Particular care must be taken if the window in the housing becomes fogged
over or discolored because moisture, water or product might seep through the
wire sheath into the terminal compartment in the housing.

Warning
Electromagnetic compatibility is only achieved if the electronics housing is
closed. Leaving the enclosure open can lead to electromagnetic disturbances.

CMB & CT OPERATING MANUAL
Part No 06EN003384 rev. 1.6

Page 35 of 106