Temperature unit, Temperature lower-range value, Temperature upper-range value – Fluid Components International CMB User Manual

Page 73

Advertising
background image

13.7.1 Temperature

unit


This function allows the operator to set the unit for temperature measurement. Press

↵ to display the

following selection field:

Temperature in

[°C]


As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can choose between
°C, °F and K. All display windows, measuring ranges and limiting values refer to the selected unit.

13.7.2 Temperature lower-range value


This function allows the operator to define the lower-range value for temperature measurement. Lower
temperatures will set the assigned current output to the minimum value of 0/4 mA. The temperature is
entered in the set temperature unit. After selecting the Temperature lower-range value function, press

to display the following selection field:

Temperature

0% = + 005 °C


The current lower-range value for temperature measurement will be displayed. As mentioned in Section
12.4.3.2 Input window/modify a value, the operator can change the value. After setting the new lower-
range value, press

↵ to confirm and apply the change.


13.7.3 Temperature upper-range value


This function allows the operator to define the upper-range value for temperature measurement. For this
temperature, the assigned current output will be set to the upper-range value of 20 mA. The applied cur-
rent of the current output assigned to the temperature value is linearly interpolated based on the ratio of
the measured value to the difference between lower- and upper-range value.

The temperature is entered in the set temperature unit. After selecting the Temperature upper-range
value

function, press

↵ to display the following selection field:

Temperature

100 % = +090 °C


The current upper-range value for temperature measurement will be displayed. As mentioned in Section
12.4.3.2 Input window/modify a value, the operator can change the value. After setting the new upper-
range value, press

↵ to confirm and apply the change.


CMB & CT OPERATING MANUAL
Part No 06EN003384 rev. 1.6

Page 73 of 106

Advertising